The EEG performance tester is compliant with the test requirements of IEC 80601-2-26:2019 and applicable for R&D, compliance and production line testing.
Parameter | Specification |
---|---|
Main output voltage accuracy |
±1% for amplitudes of 50 µVpp or higher |
Main output voltage resolution (DAC resolution) |
0.5µV |
Frequency / pulse repetition rate accuracy |
±1% |
Pulse duration / timing accuracy |
±1ms |
Resistor tolerance |
±1% |
Capacitor tolerance |
±5% |
Precision 500:1 divider (50KΩ:100Ω ) |
±0.2% |
Sample rate |
5kHz ± 0.1% |
DC offset (fixed, noise free, from internal super capacitor) |
150mV ± 1% |
DC offset (variable, may include up to 50µVpp noise) |
Setting ±1% or ±3mV |
Power supply |
USB +5Vdc supply (no separate power supply required) 0.5A (high power mode) |
Environment |
15 ~ 30°C (by design, not tested) 30 ~ 80% RH(design not tested) |
Safety, EMC standards |
No applicable safety standards (maximum internal voltages 12Vdc) For EMC no testing performed. CE marking based on careful selection of parts, including USB protection IC, as well as special filters to reduce noise from microprocessor (8MHz) and DC/DC converter (200kHz). |
SEEG 100E Test System | |
---|---|
Model No. | Description |
SEEG 100E | Single Channel EEG Test System with one EEG Breakout Box and Assistant Software for IEC 80601-2-26:2019 performance tests. |
Optional Accessory | |
---|---|
Model No. | Description |
WUI100 | USB isolator for reducing the power noise from PC. Recommanded to use with SECG 4.0, MECG 2.0, HRS200, HRS100+, SEEG 100 and SEEG 100E. |
Optional Calibration Service and Warranty Extension | |
---|---|
Model No. | Description |
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 | Extend the limited warranty from (1) year to (3) years. |