Designed specifically for AED/Defibrillator production line and QA applications to solve the need to integrate most devices. Withstand more intense defibrillation shocks at a long duration. Display defibrillation shock waveform, energy (in Joules), parameters, as well as characteristic points at the same time. Supporting various ECG waveforms with customized user-defined data or clinical signals. Also, SDK provided to enhance AED manufacturing automation.
*Note: Optional accessories or functions such as the barcode scanner, Repeatability and Reproducibility Analysis, and paddle box set need a separate purchase.
- AED Testing Scenario
- Defibrillator Testing Scenario
Energy Measurement Specifications
Items | Specifications |
---|---|
Maximum Energy |
Range: Up to 600J Accuracy: ± 1% of reading ± 1J for up to 360J; ± 1% of reading ± 1J, typical for 360J ~ 600J |
Voltage |
Range: Up to 3000V Accuracy: ± (1% of reading + 1V) for 450V ~ 3000V & -450V ~ -3000V; ± (1% of reading + 2V) for -450V ~ 450V |
Current |
Range: 1.2 ~ 100.0A Accuracy: ± (1% of reading + 0.1A) |
Load Resistance | Resistance: 50Ω |
Pulse Width |
Range: 1.0 ~ 48.0ms Accuracy: ±0.05ms |
Sample Rate |
250kHz (4 μs sample) |
Maximum Average Power |
12W, equivalent to 1 defib pulse of 360J every 30 seconds |
ECG Signal Specifications
Items | Specifications |
---|---|
Ventricular Fibrillation - Coarse | Vp-p = 3mV |
Ventricular Fibrillation - Fine |
Vp-p = 0.5mV |
Ventricular Tachycardia (≒180BPM) |
Vp-p = 3mV |
Ventricular Tachycardia Plus (≒234BPM) | Vp-p = 3mV |
Atrial Fibrillation |
Vp-p = 1.2mV |
Asystole |
Vp-p = 0.03mV |
User-defined Pattern (Max. 5K Sample Rate) |
Up to 10 sets with 100 seconds max. each set |
General Specifications
Items | Specifications |
---|---|
Temperature |
Operating: 0°C ~ 50°C (32°F ~ 122°F) Storage: -20°C ~ +60°C (-4°F ~ +140°F) |
Humidity | 10% ~ 90% non-condensing |
Communications |
USB Type-C connector |
Operation Modes | Remote (PC Software via API and SDK) |
Power |
USB 3.0 Bus Power |
Mechanical |
Housing: ABS Plastic Size (L x W x H): 19.0 x 12.5 x 6.0cm |
Compliance Specifications
Items | Specifications |
---|---|
Safety Standards | CE: IEC / EN 61010-1:2010 + A1:2016; Pollution degree 2 61010-2-030:2017 |
Electromagnetic Compatibility Standards (EMC) |
CE: EN61326 -1 FCC: EMC P15B |
Specifications are subject to change without prior notice.
DFS360 Test System
Model No. | Image | Description |
---|---|---|
300-DFS36S |
|
DFS360 AED/Defibrillator Tester Main Unit |
K22-0500101 |
|
Open-ended AED Cable (Banana Male), 0.5 meter (Black) |
K22-0500102 |
|
Open-ended AED Cable (Banana Male), 0.5 meter (Red) |
N61-0210033 |
Compact Splicing Connector |
|
K27-1800304 |
USB Transfer Cable: Type-A to Type-C (Male to Male), 1.8 meter |
Optional Paddle Box Set
Model No. | Image | Description |
---|---|---|
300-DFPDOCK |
|
DFS Tester Docking Base for Defibrillator |
E13-070P101 |
|
DFS Tester Defibrillation Receiving Plate |
K22-0100301 | ![]() |
AED Cable for Paddle Box, Female to Male (Black) |
K22-0100201 | ![]() ![]() |
AED Cable for Paddle Box, Female to Male (Red) |
Optional Tool
Model No. | Image | Description |
---|---|---|
800-DF07001 |
|
Scan Software and Barcode Scanner (Argox barcode scanner, model no. AS-9400) |
Optional Software Add-On Pack
Model No. | Description |
---|---|
HE0-DF00001 |
Repeatability and Reproducibility Analysis for inspecting DUT's quality and performance |
Optional Calibration Service and Warranty Extension
Model No. | Description |
---|---|
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 |
Extends the limited warranty from (1) year to (3) years. |