The AED inspection tester easily and quickly verifies whether the installed AED is functioning normally. It is suitable for all AED brands on the market and is dedicated to after-sales service. It generates three kinds of ECG signals, VT, VF, and NSR. In addition, it can receive AED defibrillation shocks. The user may define pass-criteria of AED shock energy to ensure that the AED works in the normal range.
Note: The DFS100 PC Software is solely for customized setting. It does not include testing functions.
Energy Measurement Specifications
Items | Specifications |
---|---|
Maximum Energy | Range: Up to 600J |
Voltage |
Range: Up to 5000Vp Accuracy: ± (1% of reading +2V) |
Load Resistance | Resistance: 50Ω |
Maximum Average Power | 12 W, equivalent to 1 defib pulse of 360 J every 30 seconds |
Self-defined pass criteria range |
40 ~ 360J |
ECG Signal Specifications
Items | Specifications |
---|---|
Waves |
Ventricular Fibrillation, Ventricular Tachycardia (≥ 234bpm) |
Normal Sinus Rhythm (NSR) |
Set at 80BPM, 3mV |
General Specifications
Items | Specifications |
---|---|
Temperature |
Operating: 0°C ~ 50°C (32°F ~ 122°F) Storage: -20°C ~ +60°C (-4°F ~ +140°F) |
Humidity | 10% ~ 90% non-condensing |
Communications |
USB Type-C connector only |
Power |
9V alkaline battery * 1 (no charger) |
Mechanical |
Housing: ABS Plastic Size (L x W x H): 17.0 x 8.55 x 4.0cm Weight: 330g |
Compliance Specifications
Items | Specifications |
---|---|
Safety Standards |
CE: IEC / EN61010-1:2010 + A1:2016; Pollution degree 2 61010-2-030:2017 |
Electromagnetic Compatibility Standards (EMC) |
CE: EN61326-1; EN301489‐1/EN301489‐17 FCC: EMC P15B |
Specifications are subject to change without prior notice.
DFS100 Test System
* Customization of AED interface cable adapter is available. Please contact WhaleTeq (service@whaleteq.com)
Optional Accessories
Model No. | Image | Description |
---|---|---|
K27-1800304 |
USB transfer cable: Type-A to Type-C (Male to Male), 1.8 meter |
|
G34-0600101 | ![]() |
DFS carrying bag |
Optional Calibration Service and Warranty Extension
Model No. | Description |
---|---|
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 |
Extend the limited warranty from (1) year to (3) years. |