The first parallel testing ECG tester designed for production lines can simultaneously test ECG function on up to 4 wearable devices with a single computer to accelerate testing time. This tester provides IEC 60601-2-47 test items and outputs 7 types of standard ECG waveforms to assist DUTs in conforming with regulatory requirements. For customized testing, the raw data playback function can reproduce clinical data or user-defined waveforms to help users verify that the DUTs meet the manufacturer's specifications. This tester integrates needed equipment to optimize testing efficiency while ensuring product quality and a stable production process. It’s suitable for manufacturers who have mass-production needs on wearable devices with the ECG function. Compliance tests conforming with IEC 60601-2-47, China YY0885, and YY 9706.247-2021

Overview
Specification
Ordering information
  • Integrates IEC 60601-2-47 test items*, including dynamic range, input impedance, frequency response, etc.
  • Equipped with 4 sets of electrode terminals - RA/LA/RLD electrode terminals on each set to meet testing needs for wearable device's ECG function
  • Outputs 7 types of standard ECG waveforms - ECG Waveform, Sine, Triangle, Square, Exponential, Rectangle Pulse, and Triangle Pulse
  • Provides SDK - users can flexibly develop automated testing programs and control multiple WECG400 to test more than 4 DUTs simultaneously
  • Supports sequential testing - tests DUTs one at a time according to a user-defined sequence
  • Equipped with 4 DUT Monitor ports to connect to an oscilloscope - amplifies the ECG signals to help users monitor the signal output with higher resolutions
  • Rack-mountable and easy to deploy with the 1U height design

*Note: For the two test items as system noise and common mode rejection, please refer to the CMRR 3.0+ Common Mode Rejection Ratio Tester for testing.

 

 

【Testing Scenario - Single WECG400】

 

【Testing Scenario - Multiple WECG400】

Standard ECG waveforms

Parameters Range Resolutions

ECG Waveform

Frequency (Hz)

0.02 ~ 6Hz

0.01

Frequency (BPM)

1 ~ 360BPM

1

Amplitude

(-5) ~ 5mV

0.01

Sine

Frequency (Hz)

0.05 ~ 500Hz

0.01

Frequency (BPM)

3 ~ 30000BPM

1

Amplitude

(-10) ~ 10mV

0.01

Triangle

Frequency (Hz)

0.05 ~ 500Hz

0.01

Frequency (BPM)

3 ~ 30000BPM

1

Amplitude

(-10) ~ 10mV

0.01

Square

Frequency (Hz)

0.05 ~ 500Hz

0.01

Frequency (BPM)

3 ~ 30000BPM

1

Amplitude

(-10) ~ 10mV

0.01

Exponential

Frequency (Hz)

0.05 ~ 5Hz

0.01

Frequency (BPM)

3 ~ 300BPM

1

Amplitude

(-5) ~ 5mV

0.01

Rectangle Pulse

Frequency (Hz)

0.05 ~ 5Hz

0.01

Frequency (BPM)

3 ~ 300BPM

1

Amplitude

(-5) ~ 5mV

0.01

Pulse width (frequency: 0.05Hz)

1 ~ 300ms

1

Pulse width (frequency: 5Hz)

1 ~ 200ms

1

Triangle Pulse

Frequency (Hz)

0.05 ~ 5Hz

0.01

Frequency (BPM)

3 ~ 300BPM

1

Amplitude

(-5) ~ 5mV

0.01

Pulse width (frequency: 0.05Hz)

1 ~ 300ms

1

Pulse width (frequency: 5Hz)

1 ~ 200ms

1

 

Input Impedance / Frequency Response Signal add-on

Parameters Range Default Values Resolutions

Play raw data

Sample rate

Maximum:

40K Hz

-

-

DC offset (non-Variable)

-300, 0, +300mV

0mV

300

DC offset (Variable)

(-1000) ~ 1000mV

0mV

100

Input impedance switch

100Ω (short circuit)

-

-

620kΩ / 4.7 nF on/off

Off

(on: short circuit)

-

10MΩ / 180 pF on/off

Off

(on: short circuit)

-

 

Product System Capacity and Standard Requirements Table

Parameters Specifications System Capability

Main output voltage accuracy

±1% for amplitudes of 0.5mVpp or higher

±0.3%

Main output voltage resolution

(DAC resolution)

N/A 0.15µV

Frequency / pulse repetition rate accuracy

±1% ±0.1%

Pulse duration / timing accuracy

±1ms ±0.2ms

Resistor tolerance

±1% ±0.5%

Capacitor tolerance

±5% ±5%

Precision 1000 : 1 divider

±0.5% ±0.3%

Sample rate

5kHz ± 0.1% 10kHz ± 0.05%(50ppm)

DC offset

(fixed, noise-free, sourced from the internal supercapacitor)

300mV ± 1% 300mV ± 0.1%

DC offset

(variable, up to 1000mV)

Setting ±1% or ±3mV

Setting ±1% or ±3mV

Environment

15 ~ 30°C (by design)
30 ~ 80% RH (by design)

5 ~ 40°C
30 ~ 80% RH
altitude < 2000M
Overvoltage CAT II

 

WECG400 Standard Package

Model No. Image Description

100-AE00009

Parallel Testing ECG Tester for wearable device production lines

 

Optional Accessories

Model No. Image Description

S92-0204021

DC12V Power Adaptor x 1

K21-0738301

RCA to BNC Cable x 2

K27-1800301

USB Cable x 1

K22-1050101

Signal Grounding Wire x 1

K29-1000801

Grounding Wire x 1

N54-0120013

Compound Terminal x 12

 

Optional Calibration Service and Warranty Extension

Model No. Description
C3 Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval.
R3

Extends the limited warranty from (1) year to (3) years.