The first parallel testing ECG tester designed for production lines can simultaneously test ECG function on up to 4 wearable devices with a single computer to accelerate testing time. This tester provides IEC 60601-2-47 test items and outputs 7 types of standard ECG waveforms to assist DUTs in conforming with regulatory requirements. For customized testing, the raw data playback function can reproduce clinical data or user-defined waveforms to help users verify that the DUTs meet the manufacturer's specifications. This tester integrates needed equipment to optimize testing efficiency while ensuring product quality and a stable production process. It’s suitable for manufacturers who have mass-production needs on wearable devices with the ECG function. Compliance tests conforming with IEC 60601-2-47, China YY0885, and YY 9706.247-2021
*Note: For the two test items as system noise and common mode rejection, please refer to the CMRR 3.0+ Common Mode Rejection Ratio Tester for testing.
【Testing Scenario - Single WECG400】
【Testing Scenario - Multiple WECG400】
Standard ECG waveforms
Parameters | Range | Resolutions | |
---|---|---|---|
ECG Waveform |
Frequency (Hz) |
0.02 ~ 6Hz |
0.01 |
Frequency (BPM) |
1 ~ 360BPM |
1 |
|
Amplitude |
(-5) ~ 5mV |
0.01 |
|
Sine |
Frequency (Hz) |
0.05 ~ 500Hz |
0.01 |
Frequency (BPM) |
3 ~ 30000BPM |
1 |
|
Amplitude |
(-10) ~ 10mV |
0.01 |
|
Triangle |
Frequency (Hz) |
0.05 ~ 500Hz |
0.01 |
Frequency (BPM) |
3 ~ 30000BPM |
1 |
|
Amplitude |
(-10) ~ 10mV |
0.01 |
|
Square |
Frequency (Hz) |
0.05 ~ 500Hz |
0.01 |
Frequency (BPM) |
3 ~ 30000BPM |
1 |
|
Amplitude |
(-10) ~ 10mV |
0.01 |
|
Exponential |
Frequency (Hz) |
0.05 ~ 5Hz |
0.01 |
Frequency (BPM) |
3 ~ 300BPM |
1 |
|
Amplitude |
(-5) ~ 5mV |
0.01 |
|
Rectangle Pulse |
Frequency (Hz) |
0.05 ~ 5Hz |
0.01 |
Frequency (BPM) |
3 ~ 300BPM |
1 |
|
Amplitude |
(-5) ~ 5mV |
0.01 |
|
Pulse width (frequency: 0.05Hz) |
1 ~ 300ms |
1 |
|
Pulse width (frequency: 5Hz) |
1 ~ 200ms |
1 |
|
Triangle Pulse |
Frequency (Hz) |
0.05 ~ 5Hz |
0.01 |
Frequency (BPM) |
3 ~ 300BPM |
1 |
|
Amplitude |
(-5) ~ 5mV |
0.01 |
|
Pulse width (frequency: 0.05Hz) |
1 ~ 300ms |
1 |
|
Pulse width (frequency: 5Hz) |
1 ~ 200ms |
1 |
Input Impedance / Frequency Response Signal add-on
Parameters | Range | Default Values | Resolutions | |
---|---|---|---|---|
Play raw data |
Sample rate |
Maximum: 40K Hz |
- |
- |
DC offset (non-Variable) |
-300, 0, +300mV |
0mV |
300 |
|
DC offset (Variable) |
(-1000) ~ 1000mV |
0mV |
100 |
|
Input impedance switch |
100Ω (short circuit) |
- |
- |
|
620kΩ / 4.7 nF on/off |
Off (on: short circuit) |
- |
||
10MΩ / 180 pF on/off |
Off (on: short circuit) |
- |
Product System Capacity and Standard Requirements Table
Parameters | Specifications | System Capability |
---|---|---|
Main output voltage accuracy |
±1% for amplitudes of 0.5mVpp or higher |
±0.3% |
Main output voltage resolution (DAC resolution) |
N/A | 0.15µV |
Frequency / pulse repetition rate accuracy |
±1% | ±0.1% |
Pulse duration / timing accuracy |
±1ms | ±0.2ms |
Resistor tolerance |
±1% | ±0.5% |
Capacitor tolerance |
±5% | ±5% |
Precision 1000 : 1 divider |
±0.5% | ±0.3% |
Sample rate |
5kHz ± 0.1% | 10kHz ± 0.05%(50ppm) |
DC offset (fixed, noise-free, sourced from the internal supercapacitor) |
300mV ± 1% | 300mV ± 0.1% |
DC offset (variable, up to 1000mV) |
Setting ±1% or ±3mV |
Setting ±1% or ±3mV |
Environment |
15 ~ 30°C (by design) |
5 ~ 40°C 30 ~ 80% RH altitude < 2000M Overvoltage CAT II |
WECG400 Standard Package
Model No. | Image | Description |
---|---|---|
100-AE00009 |
![]() |
Parallel Testing ECG Tester for wearable device production lines |
Optional Accessories
Model No. | Image | Description |
---|---|---|
S92-0204021 |
![]() |
DC12V Power Adaptor x 1 |
K21-0738301 |
![]() |
RCA to BNC Cable x 2 |
K27-1800301 |
![]() |
USB Cable x 1 |
K22-1050101 |
![]() |
Signal Grounding Wire x 1 |
K29-1000801 |
![]() |
Grounding Wire x 1 |
N54-0120013 |
![]() |
Compound Terminal x 12 |
Optional Calibration Service and Warranty Extension
Model No. | Description |
---|---|
C3 | Provides (3) years of calibration service coverage. WhaleTeq equipment can be calibrated to original performance on the basis of (1) year interval. |
R3 |
Extends the limited warranty from (1) year to (3) years. |