Test Solution for
IEC 60601-2-25

WhaleTeq offers ECG / EKG test solutions compliant with international medical standard of IEC60601-2-25, including SECG 5.0 AIO for performance tests, MECG 2.0 for  database playback, CMRR 3.0+ for common mode rejection ratio tests and CDCA software for CTS/CSE database compliance analysis.

 

Moreover, WhaleTeq TRF Generator software is the ultimate test solution for compliance testing - simplify all your efforts conducting testing and generating TRF report. Work with SECG 5.0 AIO, SECG 4.0, CMRR 3.0+ and CMRR 3.0. 

 

IEC60601-2-25:2011 (DIAGNOSTIC ECG)
Test Clause SECG 5.0 AIO MECG 2.0 CMRR 3.0+ CDCA HFCN
201.12.1.101.2 Requirements for amplitude measurements   O   O  
201.12.1.101.3.1 Requirements for absolute interval and wave duration measurements   O   O  
201.12.1.101.3.2 Requirements for interval measurements on biological ECGS   O   O  
201.12.4.101 Indication of inoperable ELECTROCARDIOGRAPH O        
201.12.4.102.3.2 Goldberger and Wilson LEADS O O      
201.12.4.102.4 Recovery time O        
201.12.4.103 Input impedance O        
201.12.4.105.1 COMMON MODE REJECTION     O    
201.12.4.105.2 Overload tolerance O        
201.12.4.105.3 FILTERS (including line frequency interference FILTERS)   O      
201.12.4.106.1 NOISE level     O    
201.12.4.106.2 CHANNEL crosstalk O        
201.12.4.107.1.1.1 High frequency response O        
201.12.4.107.1.1.2 Low frequency (impulse) response O        
201.12.4.107.1.2 Test with calibration ECGS O O      
201.12.4.107.2 Linearity and dynamic range O        
201.12.4.108.3.2 Recording speed O        
201.12.4.109 Use with cardiac pacemakers O        
202.6.2.101 Eelectrosurgery interference         O